8 Troubleshooting
80 R-220 SE Operation Manual, Version F
8.2 Error messages and their remedy
The error messages are shown in the rotational speed display.
Errors which are not listed in the table are to be corrected by a BUCHI trained technician who has
access to official service manuals. In this case, please refer to your local BUCHI customer service
agent.
Error number
Error group
“ERROR” Prefix
Errors and their remedy
Error Additional error indicator on display Possible cause Remedy
E01 Heating bath icon flashes. Bath temp. sensor not connected or
defective.
Switch off the unit. Wait for the bath
to cool down a little before restart.
Call service if problem is persistent.
E02 Heating bath icon flashes. The bath does not reach the
intended maximum height.
Lift motor is defective or
blocked.
Press the “Aerate” button to confirm
the error message.
Switch off the unit. Check if lift is
blocked.
Call service if problem is persistent.