5 Putting into operation
56 R-220 SE Operation Manual, Version F
5.15 Heating bath setup for 2×1800 W heating element
The heating element holder is labelled with the heating power data. The heating bath of the 2×1800 W
version can be filled with different kinds of liquid heating transfer medium. Depending on the
maximum heating bath temperature for the target application, a suitable medium must be used.
How to set the maximum setpoint temperature
Limiting the maximal temperature set-point to a value below 180 °C can help to eliminate processing
errors, e.g. when working with heat-sensitive substances.
AERATE
Temp
+
Set Temp
UP
DOWN
UP
a
b
Store setting
Keep the "TEMP UP" button pressed while switching on
the Rotavapor to enter the setup menu.
Set the temperature limit via the "TEMP UP" and "TEMP
DOWN" buttons. The set value is displayed as bath
temperature value on the display.
Press "AERATE" button to store the setting and to
leave the menu.