OperationManualRotavapor®R100 37/60
6.3 Finishingdistillation
u Raisetherotarydriveunit(seeChapter6.2.1"Raisingandloweringtheevaporat
ingflask",page33).
u VenttheRotavapor.
u Turnofftherotarydriveunit(settherotationspeedto0rpm).
u Switchofftheheatingbath
u CAUTION!Riskofskinburnsfromhotevaporatingflask.Checkthetempera
tureoftheevaporatingflaskand,ifnecessary,allowittocooldownorelsewear
suitableprotectivegloves.
u Removetheevaporatingflask(seeChapter6.3.1"Removingtheevaporating
flask",page37).
u Turnoffthecoolantsupply.
u Removethereceivingflask(seeChapter6.3.2"Removingthereceivingflask",
page38).
u DrytheRotavapor(seeChapter7.6"Removingsolventaccumulations",page41).
u CleantheRotavaporandallglasscomponents(seeChapter7"Cleaningandser
vicing",page39).
6.3.1 Removingtheevaporatingflask
u Raisetherotarydriveunit(seeChapter6.2.1"Raisingandloweringtheevaporat
ingflask",page33).
u VenttheRotavapor.
u Turnofftherotarydriveunit(settherotationspeedto0rpm).
u CAUTION!Riskofskinburnsfromhotevaporatingflask.Checkthetempera
tureoftheevaporatingflaskand,ifnecessary,allowittocooldownorelsewear
suitableprotectivegloves.
u Firmlyholdtheevaporatingflask(4)andreleasethecombiclip(3)byscrewingit
counterclockwise.
u Pushtheclip(2)backfromtheneckoftheevaporatingflask.
u Detachtheevaporatingflask.
u Iftheevaporatingflasksticksbecauseofchemicalresidues,pressthelockrelease
button(1)andtwisttheflask(4)counterclockwisetoremoveit.