BÜCHILabortechnikAG Operation|6
OperationManualRotavapor®R100 35/60
6.2.3 Optimizationdistillation
Dependingonthesolventinvolved,itmayalsobepossibletofurtheroptimizethedis
tillationprocess.Thedistillationshouldalwaysbesetupsothatthecondensatedoes
notcovermorethanthreequartersofthecoilinthecondenser.Thetopquartershould
remainexposed.
WARNING
Riskofexplosionfromexcessiveinternalpressure
Theevaporatingflaskand/orthecondensercouldexplodeiftheinternalpressurepro
ducedbyevaporationbecomestoogreat.
u Makesurethattheinternalpressureinsidetheglasscomponentsisnevergreater
thanatmosphericpressure.
Ifthecondensationisbelow3/4ofthecondenser:
u increasethevacuum.
Thatwilllowertheboilingpointandmorevaporwillenterthecondenser.
u Inapplicationswherenovacuumisapplied,raisethetemperatureoftheheating
bathifnecessary.
Thatwillcausemoresolventtoevaporate.
Ifthecondensationisabove3/4ofthecondenser:
u decreasethevacuum.
Thatwillraisetheboilingpointandlessvaporwillenterthecondenser.