BÜCHILabortechnikAG Installation|5
PrepChromC700 41/110
5.2 Tubing,installation
5.2.1 6wayelectricalvalve,overview
Injection adapter
Pre-Column
Column
Waste
Mixing chamber
Fraction collector
or loop
Fig.16:Overview6wayelectricalvalveconnection
5.2.2 Tubinggeneral,installation
u Installthetube(2)betweenthemixingchamber(4)andthe6wayelectri
calvalve(1),portno4.
HAF780TubingPEEK,1.6mmID
u Tightenthefittingonthe6wayelectricalvalve.
Wrench3/8"
NOTE
Turnthepiston(3)ofthemixingchambersothatthemixingrateonthepiston
canberead.
u Tightenthefittingonthemixingchamberandcounterholdthepiston(3).
Wrench3/8"
Wrench11/32"