5 Putting into operation
52
K-375/376/377 Operation Manual, Version B
5.6 Storage tank connection
To connect the storage tanks, proceed as follows:
Cut the solaflex tube into pieces to the appropriate length.
Insert a PTFE suction tube into the solaflex tube.
Push a EPDM sealing ring over the solaflex tube.
Now fasten the tubes to the tank with the red screw cover.
The storage tanks should not be positioned higher than the device itself and not lower than
1 meter below the instrument.
Fig. 5.7: Tank connection
All pumps are self-priming, no overpressure is necessary at the tanks.
NOTICE
Risk of device damage by calciferous water or wrong rong connected tanks.
Use only distilled water for the H2O storage tank to keep the steam generator maintenance-
free.
Make sure that the tanks are connected correctly. If the wrong tank (e. g. reagent tank
containing NaOH) is connected to the pump labelled as „H2O“, the steam generator will get
damaged.