7 Maintenance
132 K-375/376/377 Operation Manual, Version B
Open the service door.
Disconnect the water connection ①.
Disconnect the NaOH connection ②.
Remove the sample tube and the dip
tube ①.
Remove the screw cap with the sealing
as described in section 7.4.1 .
Loosen the screw cap with the open end
spanner (11058252) ann slide it back .
Loosen the two screws and take the
holder away.
Take the Splash protector away from the
instrument and replace it with a new one.
Reassemble in reverse order.