32/42 OperationManualInterfaceI100
8.2 Errormessages
Errormessage Remedy
Pressuresensorisdefective. u ContactBUCHICustomerService.
Pressuresensornotcalibrated. u Calibratepressuresensor(seeChap
ter6.3"Calibratingthepressuresen
sor",page26).
DatatransmissionerroronRS485. u CheckRS485leadandmakesure
thatitisonlyconnectinganinterface
unittoarecirculatingchiller.
Excesspressureinsystem. u Setpumptocontinuousmodeuntil
thesystempressureisreduced.
Errorreadingfromorwritingtomemory
(EEPROM).
u ContactBUCHICustomerService.
8.3 Customerservice
Repairstothedevicemayonlybecarriedoutbyauthorizedservicetechnicians.The
servicetechnicianshavebeencomprehensivelytechnicallytrainedandareawareof
thepotentialhazardsthatcanarisefromthedevice.
TheaddressesoftheofficialBUCHICustomerServiceofficescanbefoundonthe
BUCHIwebsiteat:www.buchi.com.Ifyouhaveanyquestionsregardingtechnicalis
suesorfaults,pleasecontactthoseoffices.
CustomerServicescanofferyou:
supplyofspareparts
repairs
technicaladvice