BÜCHILabortechnikAG Helpwithfaults|8
OperationManualInterfaceI100 31/42
8 Helpwithfaults
8.1 Faults,possiblecausesandremedies
Fault Possiblecause Remedy
Devicedoesnot
work
Deviceisnotconnectedto
powersupply
u Checkpowersupply(see
Chapter5.2"Connections",
page18).
Valveorpump
frequently
switchesover
Systemleaking u Checksystemforleaks(see
Rotavaporoperatinginstruc
tions).
u Ifnecessary,replacetubing
and/orseals.
Hysteresissettingtoosmall u Selectalargerhysteresis
(forterminalvacuums
greaterthan700mbar,setto
automatichysteresis;see
Chapter6.2.1"Manual
mode",page23).
Valvedoesnot
switchover
Valvecoildirty u ContactBUCHICustomer
Service.
Valveleadnotpluggedin u Pluginvalvelead.
Valvenotadjusted u ContactBUCHICustomer
Service.
Vacuumisnot
achieved
Backevaporationfromthere
ceivingflask
u Emptyreceivingflask(see
Rotavaporoperatinginstruc
tions).
Systemleaking u Checksystemforleaks(see
Rotavaporoperatinginstruc
tions).
u Ifnecessary,replacetubing
and/orseals.
Waterpressureofglassfilter
pumptoolow
u Increasewaterflow(see
pumpoperatinginstructions).
Vacuumpumptooweak u Usesuitablydimensioned
vacuumpump.