4 Description of function
25 R II Operation Manual, Version E
4.3 Heating bath
The heating bath can be used as a water heating bath up to 95 °C.
1
2
3
4
5
a Power supply of heating bath
b Mains switch (splash water protected)
c Adjusting knob for setting the bath tempera-
ture
d Heating LED
e Handle for easy transport and movement
Fig. 4.5: Heating bath next to the base plate
4.4 Combi-Clip
Fig. 4.6: Combi-Clip
Depending on the direction the patented
Combi-Clip is turned, it performs the following
functions:
Handling the evaporating flask:
• Release the evaporating flask by turning the
clip counterclockwise.
NOTE
Secure the flask with one hand so that it does
not fall down.
• Fix the mounted evaporating flask by
pressing down the clamp and turning the
clip clockwise.
Handling the vapor duct:
• Remove the evaporating flask first, then turn
the clip clockwise until the vapor duct is
released.
NOTE
Press the lock button to block the drive unit
(position 5 in Fig. 4.2) while removing the
vapor duct or fixing the evaporating flask.