6 Operation
28 K-425 / K-436 Operation Manual, Version A
!
WARNING
Death or serious poisoning by contact or incorporation of harmful substances at use.
Before operation, check the instrument for correct assembling•
Close unused suction module positions with glass caps•
Rinse suction modules, hoses and connectors before use•
Before operation inspect sealings and tubes for good condition and tightness•
Exchange worn out or defective parts immediately•
Only operate the instrument in a fume hood•
Directly withdraw released gases and gaseous substances by a Scrubber or a water jet pump•
Dispose of condensate in tubings and glassware safely after use•
!
CAUTION
Risk of minor or moderate cuts when handling damaged glass parts.
Handle glass parts with care•
Do not heat up empty sample tubes in the device•
Visually inspect every glass part before mounting•
Exchange damaged glass parts immediately•
Do not touch cracks or bits of broken glass with bare hands•
6.2.1 Kjeldahl digestion —preparing sample tubes
During the preparation process, handling with strong acid and other dangerous substances is inevi-
table. Thus, all steps must be performed under safe laboratory conditions!
Prerequisites
All sample tubes must be clean and free of damages.•
To achieve a reliable and reproducible digestion result the quality of sample material and additives •
(e.g. acid and catalyst) must be free of impurities and of adequate amount.
Filling of a sample tube
Take the sample tubes sequentially and weigh in sample material.•
Add calatyst to the samples (e.g. Kjeldahl tablets, order no. 028765).
Carefully(!)• add sulfuric acid (≈ 98 %) of sufficient amount to the samples.
The sample tubes can now be installed in a rack.
Note
Note down the sample weight and position to keep the probe processing and its analysis repro-•
ducible. All racks are marked on top with a "1" for the first sample position.
The rack(s) can be used as a temporary storage for empty tubes during the preparation process.•
When adding the sulfuric acid, rotate the tube to wash down any sticky sample material to the •
bottom and to coat the inside of the tube.