5 Putting into operation
28 K-415 Operation Manual, Version B
5.3 Equipping the Scrubber with the condenser unit
To prepare a Scrubber model for operation with a condenser, proceed as follows:
Unless the carrier-plate a for the condenser is not yet installed, slide it into the housing from the
side and fix it with the two screws from below. The groove b must point away from the opening in
the housing (see figure 5.5).
Fig. 5.5: Fixing the carrier plate
Carefully slide the condenser from top into the opening in the housing and fix it on the carrier plate
a with the rubber strap c. The rubber strap has to be passed through the groove on the back-
side of the carrier plate to prevent it from sliding up or down and to guarantee a tight positioning of
the condenser (see arrow in figure 5.6). Make sure the connections of the condenser are pointing
backward.
The receiving vessel d can be attached from below and secured with the catchpot clip e.
Fig. 5.6: Fixing the condenser with the rubber strap