Buchi KjelDigester K-449 Manuel d'utilisateur Page 37

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 74
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 36
6 – 37Operation Manual KjelDigester K-446/K-449
6 Operation
6.3.2 K-449
1. Press Method [METH] and select method number 0 by pressing c[+/-], pressd
[ENTER] to confirm.
2. Method number 0 is used for a manual digestion or to heat the device without lift
movement.
3. Pressa [TEMP] and adjust the temperature by pressingc [+/-].
4. Pressd [ENTER] to confirm the new value.
5. Pressb [TIME] and adjust the time by pressingc [+/-].
6. Pressd [ENTER] to confirm the new value.
6.4 Editing parameters during a digestion
NOTE
Parameters on the K-449 cannot be changed during a digestion in the automatic mode
(METH1 – 9).
1. Pressa [TEMP] and adjust the temperature by pressingc [+/-].
2. Pressd [ENTER] to confirm the new value.
3. Pressb [TIME] and adjust the time by pressingc [+/-].
4. Pressd [ENTER] to confirm the new value.
Vue de la page 36
1 2 ... 32 33 34 35 36 37 38 39 40 41 42 ... 73 74

Commentaires sur ces manuels

Pas de commentaire