3 Technical data
15 K-350 Operation Manual, Version F
3.2 Technical data
Table 3-4: Technical data K-350
Distillation Unit K-350
Power consumption max. 2.2 kW
Connection voltage 230 V ± 10% / 200 V ± 10%
Frequency 50/60 Hz
Current consumption (230 V) 8.5 A
Mains connection 3-pole (P, N, E) via power cord
Recovery rate 99.5%
Reproducibility (RSD) 1%
Detection limit 0.1 mg Nitrogen
Environmental conditions
Temperature
Altitude
Humidity
for indoor use only
5 - 35 °C
up to 2000 m
maximum relative humidity 80% for temperatures up to
31°C decreasing linearly to 67% relative humidity at 35 °C
Installation category II
Pollution degree 2
Dimensions (W x H x D) 400 x 660 x 360 mm
Weight 16.5 kg
3.3 Materials used
Table 3-5: Materials used for the K-350
Component Material designation Material code
Housing Polyurethane PUR / UL V0
Glass parts Borosilicate glass 3.3 DIN/ISO 3585
Steam generator isolation Ceramic fiber Multitherm 550
Steam generator housing Stainless steel 1.4301
Protective door Polymethyl methacrylate PMMA
Condenser cover Polymethyl methacrylate PMMA
Connecting stopper Hypalon CSM
NaOH tube Ethylene Propylene Dimonomer EPDM