C-650 Operation Manual, Version B
5 Putting into operation
34
Factory Menu Code Screen
Factory Menu Code ____
Authorized persons only
Fig 5.25: Factory Menu Code screen
The Factory Menu Code screen is used by the service engineer to access a variety of service
features and is not designed to be employed by the user.
5.9 Initial test procedures
5.9.1 Preliminary activities
This section presents a protocol that can be used to ensure that the detector is working in the
proper way.
When the device is set-up, the procedures indicated below should be performed to determine
the specific characteristics of your unit. We suggest that you save the results in a permanent
location, as they can be very useful when you are performing troubleshooting activities.
Before starting the tests for a new device or after storage, flush the detector with water at a flow
of 1mL/min for at least 15minutes.
The following activities should be performed:
Power up the device. When the detector is shipped from factory, the gain is set to 1 and the
offset to 0mV. The Signal screen should indicate 000 (or a very small signal).
Access the Temperature/Gain screen, set the temperature to 50 °C and press . View the
Status screen and verify that the temperature is rising to the setpoint on the Status screen. Tem-
perature regulation is stable when the Temperature value stops blinking.